GSO ASTM F2504:2021
ASTM F2504:05(2014)
Gulf Standard
Current Edition
·
Approved on
01 July 2021
Standard Practice for Describing System Output of Implantable Middle Ear Hearing Device
GSO ASTM F2504:2021 Files
English
6 Pages
Current Edition
Reference Language
GSO ASTM F2504:2021 Scope
1.1 This practice defines means for describing system performance (ex vivo) and, in particular, system output of an
implantable middle ear hearing device (IMEHD) by measuring
a physical quantity that is relevant to the insertion gain and
output level of the IMEHD when implanted in the patient.
1.2 This practice is similar to headphone calibration on an
artificial ear in which the sound pressure level (in decibel
sound pressure level (SPL)) measured in the artificial ear can
be converted to patient hearing level (in decibel hearing level
(HL)) using a known transfer function, as defined by ANSI 3.7.
These measurements can then be used to predict system
parameters relevant for patient benefit such as functional gain,
maximum output, and variability. Measurements defined in this
practice should be useful for patients, clinicians,
manufacturers, investigators, and regulatory agencies in making comparative evaluations of IMEHDs.
1.3 The values given in SI units are to be considered the
standard.
Best Sellers From Health Sector
GSO 2017:2023
Gulf Standard
Personal Care and Baby Wet Wipes
GSO 2241:2023
Gulf Standard
Cosmetic Products - Body Cleaners
GSO 2063:2023
Gulf Standard
Cosmetic Products - Hair Conditioner (Balsam)
GSO 1223:2023
Gulf Standard
Cosmetic Products-Deodorants
Recently Published from Health Sector
GSO ISO 20776-2:2024
ISO 20776-2:2021
Gulf Standard
Clinical laboratory testing and in vitro diagnostic test systems
— Susceptibility testing of infectious agents and evaluation of performance of antimicrobial susceptibility test devices
— Part 2: Evaluation of performance of antimicrobial susceptibility test devices against reference broth micro-dilution
GSO ISO 27186:2024
ISO 27186:2020
Gulf Standard
Active implantable medical devices — Four-pole connector system for implantable cardiac rhythm management devices — Dimensional and test requirements
GSO ISO 28620:2024
ISO 28620:2020
Gulf Standard
Medical devices
— Non-electrically driven portable infusion devices
GSO ISO 20776-1:2024
ISO 20776-1:2019
Gulf Standard
Susceptibility testing of infectious agents and evaluation of performance of antimicrobial susceptibility test devices
— Part 1: Broth micro-dilution reference method for testing the in vitro activity of antimicrobial agents against rapidly growing aerobic bacteria involved in infectious diseases